• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EON MINI; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EON MINI; SCS IPG Back to Search Results
Model Number 3788
Device Problem Insufficient Information (3190)
Patient Problem No Information (3190)
Event Date 09/02/2014
Event Type  Injury  
Event Description
Device 2 of 4.Reference mfr report: 1627487-2014-02679, 1627487-2014-02681, 1627487-2014-02682.
 
Manufacturer Narrative
Correction numbers: 1627487-05242011-002-r, 1627487-07262012-002-r.This ipg serial number was included in field advisories.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
plano TX
Manufacturer Contact
arnie ochoa
6901 preston rd.
plano, TX 75024
9723098090
MDR Report Key4134841
MDR Text Key4948753
Report Number1627487-2014-02680
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Not Applicable
Remedial Action Recall
Type of Report Initial
Report Date 09/02/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/01/2011
Device Model Number3788
Device Lot Number2887048
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/02/2014
Initial Date FDA Received09/25/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-12192011-003-R
Patient Sequence Number1
Treatment
IMPLANT:; SCS EXTENSION: MODEL 3383 (X2)
Patient Outcome(s) Other;
Patient Age58 YR
-
-