• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC ENT BALLOON SEEKER FRNT 6X17MM; INSTRUMENT, ENT MANUAL SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC ENT BALLOON SEEKER FRNT 6X17MM; INSTRUMENT, ENT MANUAL SURGICAL Back to Search Results
Catalog Number 1830617FRT
Device Problem Imprecision (1307)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/05/2014
Event Type  malfunction  
Event Description
A medtronic ent representative reported that, while in an ear, nose & throat (ent) procedure, the site alleged an inaccuracy when navigating the sphenoid balloon.When plugging in the instrument and navigating, the tip of the balloon appeared to be multiple centimeters deeper than it actually was in the patient.Instruments other than the balloon seekers were all accurate.No specific measurements of the alleged inaccuracy, or further details of the procedure, were provided.The surgeon acknowledged the inaccuracy and opted to continue the procedure to completion, with the use of the navigation system.There was no impact on patient outcome.
 
Manufacturer Narrative
(b)(4) 2014.A medtronic representative performed a navigation system check-out, all areas passed.System performed as intended.This is a disposable device.No parts have been returned to manufacturer for analysis.
 
Manufacturer Narrative
Expiration date, lot number and associated manufacture date provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BALLOON SEEKER FRNT 6X17MM
Type of Device
INSTRUMENT, ENT MANUAL SURGICAL
Manufacturer (Section D)
MEDTRONIC ENT
xomed headquarters
6743 southpoint drive
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC.
826 coal creek circle
louisville CO 80027 971
Manufacturer Contact
amos jarrette
826 coal creek circle
louisville, CO 80027-9710
7208902082
MDR Report Key4140170
MDR Text Key15807114
Report Number1723170-2014-01044
Device Sequence Number1
Product Code LRC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132297
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 09/05/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/03/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date12/08/2015
Device Catalogue Number1830617FRT
Device Lot Number0208437770
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received10/07/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/11/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age45 YR
Patient Weight102
-
-