BAXTER HEALTHCARE - LARGO HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C8310R
Device Problems Degraded (1153); Volume Accuracy Problem (1675)
Patient Problem No Patient Involvement (2645)
Event Date 09/08/2014
Event Type  malfunction  
Event Description
During evaluation of a returned homechoice (hc) device, a baxter technician determined the hc machine failed fluid volume accuracy testing.Device failed during evaluation, no patient involved.
 
Manufacturer Narrative
(b)(4).The device was received for evaluation.Upon completion of baxter's investigation, if additional relevant information is obtained, a supplemental report will be submitted.
 
Manufacturer Narrative
(b)(4).The device was returned to baxter, and the evaluation is complete.An external and internal inspection was performed with no issues noted.The homechoice device received a returned instrument testing evaluation (rite).This evaluation included functional and electrical testing of the device.The device passed the rite electrical testing, but failed the rite functional testing for volumetric accuracy out of limits.The piston foam was inspected and deterioration was found.A service history review was performed and revealed that deteriorated piston foam was previously replaced.Upon completion of the investigation, the reported issue was verified and the cause determined to be deteriorated piston foam.The foam was scrapped.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE PRO
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer (Section G)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer Contact
christina arnt
25212 w. illinois route 120
round lake, IL 60073
2242703198
MDR Report Key4141256
MDR Text Key4795174
Report Number1416980-2014-34472
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/08/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number5C8310R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/05/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/24/2014
Initial Date FDA Received10/03/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/19/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-