Catalog Number 5C8310R |
Device Problems
Degraded (1153); Volume Accuracy Problem (1675)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/08/2014 |
Event Type
malfunction
|
Event Description
|
During evaluation of a returned homechoice (hc) device, a baxter technician determined the hc machine failed fluid volume accuracy testing.Device failed during evaluation, no patient involved.
|
|
Manufacturer Narrative
|
(b)(4).The device was received for evaluation.Upon completion of baxter's investigation, if additional relevant information is obtained, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
(b)(4).The device was returned to baxter, and the evaluation is complete.An external and internal inspection was performed with no issues noted.The homechoice device received a returned instrument testing evaluation (rite).This evaluation included functional and electrical testing of the device.The device passed the rite electrical testing, but failed the rite functional testing for volumetric accuracy out of limits.The piston foam was inspected and deterioration was found.A service history review was performed and revealed that deteriorated piston foam was previously replaced.Upon completion of the investigation, the reported issue was verified and the cause determined to be deteriorated piston foam.The foam was scrapped.Should additional relevant information become available, a supplemental report will be submitted.
|
|