• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - LARGO HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - LARGO HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C8310R
Device Problems Degraded (1153); Volume Accuracy Problem (1675)
Patient Problem No Patient Involvement (2645)
Event Date 09/09/2014
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was returned to baxter, and the evaluation was completed.No abnormalities were identified during internal/external inspection.Temperature confirmation testing was performed and the results were found to be within specifications.Electrical returned instrument test evaluation (rite) was conducted and the device passed.The failure of the fluid volume accuracy test occurred during functional rite, as fluid was transferred above the allowable specification range.A service history record review did not identify any issues in the previous service events that would have contributed to the reported issue.The cause of the volumetric accuracy failure was determined to be due to a deteriorated piston foam and cracked door piston.The piston foam and door piston was scrapped.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
During evaluation of a returned homechoice (hc) device, a baxter technician determined the hc machine failed rite (returned instrument test/evaluation) testing for fluid volume accuracy.Device failed during evaluation, no patient involved.
 
Manufacturer Narrative
(b)(4).The homechoice was received and an evaluation is being performed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE PRO
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer (Section G)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer Contact
christina arnt
25212 w. illinois route 120
round lake, IL 60073
2242703198
MDR Report Key4141315
MDR Text Key12576214
Report Number1416980-2014-34480
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/09/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number5C8310R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/28/2014
Initial Date FDA Received10/03/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/20/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-