Catalog Number 5C8310R |
Device Problems
Degraded (1153); Volume Accuracy Problem (1675)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/09/2014 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device was returned to baxter, and the evaluation was completed.No abnormalities were identified during internal/external inspection.Temperature confirmation testing was performed and the results were found to be within specifications.Electrical returned instrument test evaluation (rite) was conducted and the device passed.The failure of the fluid volume accuracy test occurred during functional rite, as fluid was transferred above the allowable specification range.A service history record review did not identify any issues in the previous service events that would have contributed to the reported issue.The cause of the volumetric accuracy failure was determined to be due to a deteriorated piston foam and cracked door piston.The piston foam and door piston was scrapped.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
During evaluation of a returned homechoice (hc) device, a baxter technician determined the hc machine failed rite (returned instrument test/evaluation) testing for fluid volume accuracy.Device failed during evaluation, no patient involved.
|
|
Manufacturer Narrative
|
(b)(4).The homechoice was received and an evaluation is being performed.Should additional relevant information become available, a supplemental report will be submitted.
|
|