While treating the patient's breast field 1 with radiation, they received a high voltage circuit breaker (hvcb) fault which interrupted treatment for this field delivering 74 out of 82 monitor units (mu's).Unable to clear the fault, the clinician got the patient off the table, cleared the fault by resetting the breaker and put the patient back on the table where they received the partial treatment message that stated "74/82 mu was delivered do you wish to continue with the treatment which has a remainder of 8 mu".The clinician selected ok and "beamed on".The clinician noticed that the beam did not stop at 8mu so she hit "beam off button" at which by this time 29 mu's was delivered to this field which gave an additional 21 mu's not planned.They spoke with their physicist and continued to treat the other fields in the plan; however, the plan is still displaying that field 1 was only partially treated with the delivered mu of 74/82.
|