The reported complaint was confirmed.The field service representative (fsr) replaced the batteries in the system-1.The fsr performed a release test and waited for the power manager to reset to 18.0 ah.The unit operated to manufacturer specifications and was returned to clinical use.The suspect batteries were returned to the manufacturer for further evaluation.Per the laboratory evaluation, the batteries were received in a severely depleted condition, the batteries measured 2.0 and 2.5 vdc.The batteries were installed on (b)(4) 2012.
|