• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KERR CORPORATION SONICFILL COMPOSITE; TOOTH SHADE RESIN MATERIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KERR CORPORATION SONICFILL COMPOSITE; TOOTH SHADE RESIN MATERIAL Back to Search Results
Device Problem Material Discolored (1170)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Event Description
A doctor alleged that five (5) patients had experienced the presence of black specks within their sonicfill restorations after light curing the composite.This is the second of five (5) reports.
 
Manufacturer Narrative
Specific patient information with regard to gender, age and weight was not provided.Although the doctor identified three (3) shades associated with the black specks, he could not verify which lot was used on each patient; therefore, no lot numbers were identified.The lots involved in the alleged incidents include lot numbers 5044580, 4819811, and 4814790.The doctor drilled out the composite and repeated the procedure during the same office visit.To date, the patient is doing fine.A visual evaluation was performed on the returned product from each of the lots, yielding results within specifications.In addition, a dhr review indicated that there were no deviations from the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SONICFILL COMPOSITE
Type of Device
TOOTH SHADE RESIN MATERIAL
Manufacturer (Section D)
KERR CORPORATION
1717 west collins avenue
orange CA 92867
Manufacturer (Section G)
KERR CORPORATION
1717 west collins avenue
orange CA 92867
Manufacturer Contact
kerri casino
1717 west collins avenue
orange, CA 92867
7145167634
MDR Report Key4153027
MDR Text Key16543381
Report Number2024312-2014-00632
Device Sequence Number1
Product Code EBF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 09/08/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Dentist
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/08/2014
Initial Date FDA Received10/08/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-