Model Number 5392 |
Device Problems
Failure to Power Up (1476); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/25/2014 |
Event Type
malfunction
|
Manufacturer Narrative
|
Further analysis was performed on the main pcb.Visual inspection was performed, no anomalies observed.Bench analysis confirmed the error number.Additional analysis indicated that the error displayed during a self-test.The error is a communication error between the microprocessor and the die stack (pacing engine).The board was x-rayed to look for any solder issues, none were observed.The die stack component was then exposed to a solder reflow process, the board still failed.A new die stack was installed on the board and it still failed.The root cause for this failure is unknown at this time.
|
|
Event Description
|
It was reported that an external pulse generator (epg) will not turn on and when it does turn on sometimes, it gives an error message.The epg was returned to the manufacturer for servicing.There was no patient involvement.
|
|
Manufacturer Narrative
|
This event occurred outside the us.Where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.Product event summary: analysis confirmed the reported event.The unit would not switch on.After reinserting the batteries, the unit starts up with an error, indicating that the main printed circuit board (pcb) was defective.The unit was then cleaned an inspected, and the main pcb was replaced.The unit was then tested and calibrated on an automated test system, then it passed testing.(b)(4).
|
|