Model Number 5392
Device Problems Failure to Power Up (1476); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 07/25/2014
Event Type  malfunction  
Manufacturer Narrative
Further analysis was performed on the main pcb.Visual inspection was performed, no anomalies observed.Bench analysis confirmed the error number.Additional analysis indicated that the error displayed during a self-test.The error is a communication error between the microprocessor and the die stack (pacing engine).The board was x-rayed to look for any solder issues, none were observed.The die stack component was then exposed to a solder reflow process, the board still failed.A new die stack was installed on the board and it still failed.The root cause for this failure is unknown at this time.
 
Event Description
It was reported that an external pulse generator (epg) will not turn on and when it does turn on sometimes, it gives an error message.The epg was returned to the manufacturer for servicing.There was no patient involvement.
 
Manufacturer Narrative
This event occurred outside the us.Where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.Product event summary: analysis confirmed the reported event.The unit would not switch on.After reinserting the batteries, the unit starts up with an error, indicating that the main printed circuit board (pcb) was defective.The unit was then cleaned an inspected, and the main pcb was replaced.The unit was then tested and calibrated on an automated test system, then it passed testing.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 1190 0
MY  11900
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4158586
MDR Text Key12883483
Report Number3004593495-2014-00035
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
K132924
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/29/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/24/2014
Initial Date FDA Received10/10/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/10/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-