Model Number DTBA1D1
Device Problem Pacing Problem (1439)
Patient Problems Bradycardia (1751); Weakness (2145); Complaint, Ill-Defined (2331)
Event Date 08/18/2014
Event Type  Injury  
Event Description
It was reported that the patient was having a slow heart rate and not feeling well.The patient was having weakness during the bradycardia episodes.It was found that the right ventricular (rv) lead was t-wave oversensing.The t-wave oversensing was interfering with the pacing timing of the device which was causing the patient to have periodic heart rates in the low fifties when the device¿s programmed lower rate was at seventy.Reprogramming was done and the device and rv lead remain in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Concomitant medical products: 694765 lead implanted: (b)(6) 2009 ; 4244 competitor lead implanted: (b)(6) 2005.(b)(4).
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was further reported that the cardiac resynchronization therapy defibrillator (crt-d) was explanted and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVA¿ XT CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
road 31, km. 24, hm 4
ceiba norte industrial park
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key4158610
MDR Text Key15814001
Report Number3004209178-2014-18946
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative,company representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/10/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/28/2015
Device Model NumberDTBA1D1
Device Catalogue NumberDTBA1D1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/04/2019
Date Device Manufactured11/11/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
419478 LEAD
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age00083 YR
Patient Weight79
-
-