Model Number 429688
Device Problems Pacing Inadequately (1442); Device Dislodged or Dislocated (2923); High Capture Threshold (3266)
Patient Problem No Code Available (3191)
Event Date 08/18/2014
Event Type  Injury  
Event Description
It was reported that due to positioning (to avoid phrenic stimulation), the lead would not stay in place.It was also reported there were high unstable thresholds.The lead was removed and replaced.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary- the full lead was returned and analyzed.No anomalies were found.The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Concomitant medical products: 1688-tc lead, (b)(6) 2004; 1668-tc lead, (b)(6) 2004; 6949-65 lead, (b)(6) 2014; 4194-88 lead, (b)(6) 2010.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN ABILITY PLUS
Type of Device
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4159482
MDR Text Key4770351
Report Number2649622-2014-10740
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P080006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 08/18/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/10/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/26/2016
Device Model Number429688
Device Catalogue Number429688
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/18/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/10/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DTBA1D4 ICD
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age00074 YR
-
-