Model Number D314TRG
Device Problems Inappropriate/Inadequate Shock/Stimulation (1574); Invalid Sensing (2293)
Patient Problem Shock (2072)
Event Date 07/02/2014
Event Type  Injury  
Event Description
It was reported that the patient received an inappropriate shock which resulted from the device detection not withholding during intermittent t-wave oversensing (twos) algorithm on the right ventricular (rv) lead.The sensitivity was reprogrammed.The device and lead remain in use.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Concomitant medical products: 429688, lead, implanted (b)(6) 2012; 407652 lead, implanted (b)(6) 2012.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTECTA XT CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM DISEASE MGMT
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
nashoane fulwood-kelley
8200 coral sea st ne
mounds view, MN 55112
7635260583
MDR Report Key4159807
MDR Text Key4868152
Report Number3004209178-2014-19189
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 07/02/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/10/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/28/2013
Device Model NumberD314TRG
Device Catalogue NumberD314TRG
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/02/2014
Date Device Manufactured12/12/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
694765 LEAD
Patient Outcome(s) Life Threatening; Required Intervention;
Patient Age00042 YR
-
-