TERUMO PHILIPPINES CORPORATION SURGUARD2 SAFETY NEEDLE; TERUMO NEOLUS NEEDLE Back to Search Results
Catalog Number SG2-2516
Device Problem Use of Device Problem (1670)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/17/2014
Event Type  malfunction  
Event Description
The user facility reported a needle stick due to safety device not engaged.Follow-up communication with the user facility confirmed; (1) the nurse was administering tdap; (2) the nurse attempted to lock the safety cap using the table top; (3) a click was herd click and device was put in the sink; (4) the nurse went back she found the device not engaged and stuck herself; (5) the nurse stated the needle was bent sideways, but was not protruding through the cap; and (5) there was no issues as a result of needle stick.
 
Manufacturer Narrative
The actual device has not been returned for evaluation and the investigation has yet to be completed.A follow-up report will be submitted when the investigation is complete.All available information has been placed on file in quality assurance for appropriate tracking, trending and follow-up.(b)(4).Device not returned.
 
Manufacturer Narrative
The involved device was not returned for evaluation.Visually inspection of retention and current lot samples did not reveal any defects or abnormalities.Function testing confirmed that performance specifications were met.A review of the device history record indicated that there were no production related problems.Although the cause of the reported event cannot be definitively determined based upon the available information, there is no evidence that this event was related to a device defect or malfunction.The device labeling does address the potential for such an occurrence in the warnings/cautions and the instructions-for-use with statements such as the following: (1) "handle with care to avoid needlesticks"; (2) "keep hands behind the needle at all times during use and disposal"; (3) "for greatest safety, activate the sheath using a one-handed technique"; (4) "if a needle is bent or damaged, no attempt should be made to straighten needle or use the product"; (5) "position sheath approximately 45 degrees to a flat surface.Press down with a firm, quick motion until a distinct audible click is heard"; (6) "visually confirm that the needle is fully engaged under the lock"; and (7) "dispose of used needles and materials following the policies and procedures of your facility as well as federal and local regulations for sharps disposal." all available information has been placed on file in quality assurance for appropriate tracking, trending and follow-up.
 
Event Description
(b)(4).
 
Manufacturer Narrative
(b)(4).
 
Event Description
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURGUARD2 SAFETY NEEDLE
Type of Device
TERUMO NEOLUS NEEDLE
Manufacturer (Section D)
TERUMO PHILIPPINES CORPORATION
124 east main avenue
laguna technopark
binan, laguna
RP 
Manufacturer (Section G)
TERUMO PHILIPPINES CORPORATION
124 east main avenue
laguna technopark
binan, laguna
RP  
Manufacturer Contact
cathleen hargreaves
950 elkton blvd.
elkton, MD 21921
8002837866
MDR Report Key4163472
MDR Text Key4785368
Report Number3003902955-2014-00020
Device Sequence Number1
Product Code MEG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051865
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional,User Facility,Company Representative
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 09/12/2014,10/10/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/10/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2018
Device Catalogue NumberSG2-2516
Device Lot Number130930B
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date07/17/2014
Device Age1 YR
Event Location Hospital
Date Report to Manufacturer09/12/2014
Date Manufacturer Received10/23/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/30/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-