• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - LARGO HOMECHOICE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - LARGO HOMECHOICE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C4471R
Device Problems Degraded (1153); Volume Accuracy Problem (1675)
Patient Problem No Patient Involvement (2645)
Event Date 09/18/2014
Event Type  malfunction  
Event Description
During evaluation of a returned homechoice (hc) device, a baxter technician determined the hc machine failed fluid volume accuracy testing.Device failed during evaluation, no patient involved.
 
Manufacturer Narrative
(b)(4).The device has been received and the evaluation is in progress.Should additional relevant information become available, a follow-up report will be submitted.
 
Manufacturer Narrative
(b)(4).Evaluation summary: the evaluation of the returned device is complete.An internal and external inspection was performed with no abnormalities noted.The homechoice device received a returned instrument testing evaluation (rite). this evaluation included functional and electrical testing of the device.The failed fluid volume accuracy test occurred during functional rite.An event history log review was performed and showed no keystrokes, programming, or use related events that indicated and/or contributed to the issue. a service history review was conducted and there were no deviations noted during the service of this device that could have caused or contributed to the issue.The cause of the functional rite test failure was determined to be a deteriorated door piston foam and a cracked door piston.The device will be serviced to meet specifications.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer (Section G)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer Contact
christina arnt
25212 w. illinois route 120
round lake, IL 60073
2242703198
MDR Report Key4167185
MDR Text Key4874798
Report Number1416980-2014-35657
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/18/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number5C4471R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received10/24/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-