Catalog Number 5C4471R |
Device Problems
Degraded (1153); Volume Accuracy Problem (1675)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/18/2014 |
Event Type
malfunction
|
Event Description
|
During evaluation of a returned homechoice (hc) device, a baxter technician determined the hc machine failed fluid volume accuracy testing.Device failed during evaluation, no patient involved.
|
|
Manufacturer Narrative
|
(b)(4).The device has been received and the evaluation is in progress.Should additional relevant information become available, a follow-up report will be submitted.
|
|
Manufacturer Narrative
|
(b)(4).Evaluation summary: the evaluation of the returned device is complete.An internal and external inspection was performed with no abnormalities noted.The homechoice device received a returned instrument testing evaluation (rite). this evaluation included functional and electrical testing of the device.The failed fluid volume accuracy test occurred during functional rite.An event history log review was performed and showed no keystrokes, programming, or use related events that indicated and/or contributed to the issue. a service history review was conducted and there were no deviations noted during the service of this device that could have caused or contributed to the issue.The cause of the functional rite test failure was determined to be a deteriorated door piston foam and a cracked door piston.The device will be serviced to meet specifications.Should additional relevant information become available, a supplemental report will be submitted.
|
|