• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC INFUSE BONE GRAFT; FILLER, RECOMBINANT HUMAN BONE MORPHOGENETIC PROTEIN, COLLAGEN SCAFFOLD WITH MET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC INFUSE BONE GRAFT; FILLER, RECOMBINANT HUMAN BONE MORPHOGENETIC PROTEIN, COLLAGEN SCAFFOLD WITH MET Back to Search Results
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Unspecified Infection (1930)
Event Type  Injury  
Event Description
It was reported that the patient underwent a tlif using rhbmp-2/acs.Post-op, the patient is "having some pretty intense complications and am currently waiting for the results of a second opinion mri and bone scan to see what is causing my problems.¿.
 
Manufacturer Narrative
(b)(4).Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSE BONE GRAFT
Type of Device
FILLER, RECOMBINANT HUMAN BONE MORPHOGENETIC PROTEIN, COLLAGEN SCAFFOLD WITH MET
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK
1800 pyramid place
memphis TN 38132
Manufacturer Contact
huzefa mamoola
1800 pyramid place
memphis, TN 38132
9013963133
MDR Report Key4173828
MDR Text Key5156648
Report Number1030489-2014-04036
Device Sequence Number1
Product Code NEK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P000058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 09/15/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/15/2014
Initial Date FDA Received10/15/2014
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-