• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS PERFORATOR DRIVER WITH HUDSON END; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS PERFORATOR DRIVER WITH HUDSON END; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Catalog Number CSR60
Device Problem Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
Customer reported that the reducer became unusually warm.Device was returned for repair.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Synthes is submitting this report as a result of remediation activities related to synthes ous service and repair files legacy review/remediation protocol-complaint handling and mdr reporting.Device listed in this report is used for treatment, not diagnosis.Any additional information received regarding this event after filing this report shall be filed on a supplemental mdr.(b)(6).During pre-repair assessment performed by a technician, functional testing was performed and revealed the gears had severe scratches, the drive shaft end play was veiled, and the device was full of detergent and rust.This was attributed to improper maintenance.The device was repaired and returned to the customer on (b)(4) 2013.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFORATOR DRIVER WITH HUDSON END
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key4175046
MDR Text Key5163001
Report Number1045834-2014-13515
Device Sequence Number1
Product Code HBE
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
PK965080
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 04/10/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCSR60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2013
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/10/2013
Initial Date FDA Received10/15/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/21/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-