HILL-ROM, INC. VERSACARE BED; A/C POWERED ADJUSTABLE HOSP BED Back to Search Results
Model Number 3200
Device Problem Material Integrity Problem (2978)
Patient Problems Wound Dehiscence (1154); Injury (2348)
Event Date 09/02/2014
Event Type  Injury  
Event Description
Hill-rom received a report from the account stating that the pt had a pre-existing stage four wound that was closed and now has reopened on the sacrum.The account also reported two new stage three wounds on the left and right ischia.The pt complained that the mattress was too hard.(b)(4).
 
Manufacturer Narrative
The hill-rom technician found error codes in the log but could not duplicate or find any issue with the bed.He tested the blower output and inspected the mattress and hose for tears, rips, and kinks and could not find any issues.He calibrated the air sys to clear the log and ease the pts concerns.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unk if the account performs preventative maintenance on their bed.The reported injury is serious in nature per fda definition.Development of pressure ulcers is multifactorial and cannot be only attributed to performance of the surface.Risk factors include protein-calorie malnutrition, microclimate (skin wetness caused by sweating or incontinence), diseases that reduce blood flow to the skin, such as arteriosclerosis, or diseases that reduce the sensation in the skin, such as paralysis or neuropathy.Position changes are key to pressure sore prevention and treatment.These changes need to be frequent, repositioning needs to avoid stress on the skin, and body positons need to minimize the risk of pressure on vulnerable areas.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACARE BED
Type of Device
A/C POWERED ADJUSTABLE HOSP BED
Manufacturer (Section D)
HILL-ROM, INC.
batesville IN
Manufacturer Contact
tony werner
1069 state route 46 east
batesville, IN 47006
8129312359
MDR Report Key4176634
MDR Text Key20453184
Report Number1824206-2014-02352
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 09/02/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Model Number3200
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/02/2014
Initial Date FDA Received09/29/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Weight49
-
-