MAUDE Adverse Event Report: VENTLAB AIRFLOW RESUSCITATION BAG; MANUAL RESUSCITATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VENTLAB AIRFLOW RESUSCITATION BAG; MANUAL RESUSCITATOR Back to Search Results
Model Number AF5140MB-9
Device Problems Failure to Deliver (2338); No Flow (2991)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/06/2014
Event Type  Injury  
Event Description
Valves won't open.Pulled out of bag and attempted to use on a patient and valve would not open.No death or serious injury reported.
 
Manufacturer Narrative
This belated mdr is being submitted as part of our remediation efforts in response to the fda form 483 notification issued on june 4, 2014 by (b)(4) to ventlab, llc.This effort consists of 48 mdr's.The affected device was not investigated.However, ventlab performed quality trending, product investigation and corrective actions to solve the root causes.The solutions were implemented last year and a recall was initiated in may of this year.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIRFLOW RESUSCITATION BAG
Type of Device
MANUAL RESUSCITATOR
Manufacturer (Section D)
VENTLAB
grand rapids MI
Manufacturer Contact
rob yamashita
2710 northridge dr. nw
ste a
grand rapids, MI 49544
6162598373
MDR Report Key4188092
MDR Text Key20065166
Report Number2246980-2014-00027
Device Sequence Number1
Product Code BTM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K012842
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date 09/19/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAF5140MB-9
Device Lot Number106907
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/10/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/06/2014
Initial Date FDA Received09/19/2014
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberRES68242
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-