• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. SJM TRIFECTA VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. SJM TRIFECTA VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number TF-21A
Device Problem Perivalvular Leak (1457)
Patient Problem No Information (3190)
Event Date 09/24/2014
Event Type  Injury  
Manufacturer Narrative
(b)(4).The results of this investigation concluded fibrous pannus ingrowth on the inflow surface of all cusps, and a thin layer of fibrin on all cusps.Special stains were negative for organisms, and no acute inflammation or significant calcifications were present.There was no evidence found to suggest the cause of the fibrin and pannus formation were due to an intrinsic defect in the valve, as supported by review of the valve's device history record and the analysis performed.The cause of the fibrin and pannus formation remains unknown.
 
Event Description
The valve was explanted due to paravalvular leakage.
 
Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM TRIFECTA VALVE
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
177 east county road b
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
177 east county road b
st. paul MN 55117
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key4200423
MDR Text Key13043673
Report Number3007113487-2014-00026
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PP100029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/24/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/07/2016
Device Model NumberTF-21A
Device Catalogue NumberTF-21A
Device Lot Number4436082
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/24/2014
Initial Date FDA Received10/24/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/02/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/08/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age82 YR
Patient Weight80
-
-