ST. JUDE MEDICAL - NEUROMODULATION OCTRODE; SCS LEAD Back to Search Results
Model Number 3186
Device Problems Kinked (1339); Device Operates Differently Than Expected (2913)
Patient Problem Inadequate Pain Relief (2388)
Event Date 09/23/2014
Event Type  Injury  
Event Description
It was reported the pt (b)(6) lost stimulation.In turn, the pt underwent surgical intervention.During the procedure the physician noted one of the pt's trial leads appeared to have an impression on the outer insulation and the lead was no longer needed.As a result, the physician explanted the trial lead which resolved the issue.
 
Manufacturer Narrative
Results: it was found that the returned lead had a kink in the outer tubing with all internal wires broken.It was concluded that the fracture in the lead was consistent with stress the lead was subjected to at the distal end of the swift-lock anchor.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
plano TX
Manufacturer Contact
benjyna obasuyi
6901 preston rd.
plano, TX 75024
9725264659
MDR Report Key4201971
MDR Text Key5151008
Report Number1627487-2014-25649
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 09/23/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/17/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/01/2015
Device Model Number3186
Device Lot Number4075000
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer10/02/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received09/23/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
IMPLANT DATE:; SCS ANCHOR, MODEL 1192,
Patient Outcome(s) Other;
-
-