ST. JUDE MEDICAL, INC., CRMD QUADRA ASSURA CRT-D, DF-4 CONNECTOR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
|
Back to Search Results |
|
Model Number CD3365-40Q |
Device Problems
High impedance (1291); Over-Sensing (1438); Decreased Sensitivity (2534)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/30/2014 |
Event Type
Injury
|
Event Description
|
It was reported that high, out of range, pacing impedance, noise, decreased r-wave amplitudes, and increased capture thresholds were observed during a clinical follow-up.The lead tested normal and the anomalous measurements were noted only when connected to the device.The device was explanted and replaced.All measurements were normal with the new device.No further issues have been reported.
|
|
Manufacturer Narrative
|
All information provided by manufacturer, no medwatch form was received.(b)(4).Evaluation description included.The reported field event of sensing and capture anomaly could not be confirmed in the laboratory.The reported field event of high, out of range pacing lead impedance was confirmed in the laboratory via programmer printouts.The device was tested on the bench and using automated testing equipment, and no anomalies were found.The cause of the high pacing lead impedance could not be determined.
|
|
Search Alerts/Recalls
|
|
|