Catalog Number 34922
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Information (3190)
Event Type  Injury  
Event Description
A doctor alleged that black specks were present in the sonicfill composite material during three (3) patients' restoration procedure.This is the second of three (3) reports.
 
Manufacturer Narrative
Specific patient information with regard to weight was not provided.The doctor drilled out the composite and repeated the procedure during the same office visit using a different product, without further incident.To date, the patient is doing fine.The product alleged in this incident was not returned; therefore, a visual inspection of the retain sample was performed, yielding results within specifications.A dhr review revealed that there were no deviations from the manufacturing process.In addition, no similar complaints were received with regard to this lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SONICFILL COMPOSITE
Type of Device
TOOTH SHADE RESIN MATERIAL
Manufacturer (Section D)
KERR CORPORATION
1717 west collins avenue
orange CA 92867
Manufacturer (Section G)
KERR CORPORATION
1717 west collins avenue
orange CA 92867
Manufacturer Contact
kerri casino
1717 west collins avenue
orange, CA 92867
7145167634
MDR Report Key4210670
MDR Text Key5148709
Report Number2024312-2014-00652
Device Sequence Number1
Product Code EBF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 09/29/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Dentist
Device Expiration Date12/01/2015
Device Catalogue Number34922
Device Lot Number5044581
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/29/2014
Initial Date FDA Received10/29/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/19/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-