• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION OCTRODE; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION OCTRODE; SCS LEAD Back to Search Results
Model Number 3186
Device Problems Kinked (1339); Device Operates Differently Than Expected (2913)
Patient Problems Scar Tissue (2060); Inadequate Pain Relief (2388); Blood Loss (2597)
Event Date 09/30/2014
Event Type  Injury  
Event Description
It was reported the patient ((b)(6)) lost stimulation.X-rays were taken and revealed a kink on the lead near the anchor site.In turn, the patient underwent surgical intervention to explant and replace the lead which resolved the issue.Additionally, it was reported the procedure was extended by approximately two hours due to scar tissue and excessive bleeding.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
plano TX
Manufacturer Contact
benjyna obasuyi
6901 preston rd.
plano, TX 75024
9725264659
MDR Report Key4212918
MDR Text Key15814558
Report Number1627487-2014-25658
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 09/30/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/24/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/01/2015
Device Model Number3186
Device Lot Number4157657
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received09/30/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
IMPLANT DATE:; SCS IPG: MODEL 3608; IMPLANT DATE:; SCS ANCHOR: MODEL 1192
Patient Outcome(s) Other;
Patient Age71 YR
Patient Weight87
-
-