Model Number CD3211-36
Device Problem False Alarm (1013)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/02/2014
Event Type  malfunction  
Event Description
It was reported that a patient notifier alerted for charge time limit reached.The measured charge time in clinic was in range and a capacitor maintenance was performed with normal charge times.A review of previous follow-ups showed that the alert occurred after the patient received several therapies in a short period.Patient condition was fine after the event.
 
Manufacturer Narrative
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMOTE PLUS CRT-D
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sharon mathew
645 almanor avenue
sunnyvale, CA 94085
4085226327
MDR Report Key4217674
MDR Text Key17274333
Report Number2938836-2014-17343
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/31/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2013
Device Model NumberCD3211-36
Device Catalogue NumberSMTFY999
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/05/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/21/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-