• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - LARGO HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - LARGO HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C8310R
Device Problems Degraded (1153); Volume Accuracy Problem (1675)
Patient Problem No Patient Involvement (2645)
Event Date 10/07/2014
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The homechoice device was returned and evaluated by the product analysis lab (pal).A review of the service history revealed no indication that the parts replaced during servicing caused or contributed to the reported event.A visual inspection was performed with no abnormalities noted.The homechoice device received a returned instrument testing evaluation (rite).This evaluation included functional and electrical testing of the device.The device passed electrical rite but failed functional rite for volumetric accuracy.The cause was determined to be a deteriorated piston foam.The piston foam was replaced to resolve the problem.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
During evaluation of a returned homechoice device, a baxter technician determined the device failed the volume accuracy test.Device failed during evaluation, no patient involved.
 
Manufacturer Narrative
(b)(4).The homechoice (hc) was returned and is currently in the process of being evaluated.Should additional relevant information become available, a follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE PRO
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer (Section G)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer Contact
christina arnt
25212 w. illinois route 120
round lake, IL 60073
2242703198
MDR Report Key4218104
MDR Text Key12695566
Report Number1416980-2014-38219
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/07/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number5C8310R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/20/2014
Initial Date FDA Received10/31/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/10/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-