Model Number 37602
Device Problem Pocket Stimulation (1463)
Patient Problems Undesired Nerve Stimulation (1980); Tingling (2171)
Event Type  malfunction  
Event Description
It was reported that the patient felt a tingling at the implantable neurostimulator (ins) site in their left chest.It was unknown if therapy had been titrated down at implant as suggested due to differences in voltage between legacy and current devices.Threshold testing had not been redone.Imaging had not shown any visible issues with the system.The issue was unresolved at the time of this report and the healthcare professional planned to address the tingling problem surgically, but they had since decided to do no intervention at the time of this report.Impedances of the left globus pallidus (gpi), right abdomen placement and right gpi, left chest subclavicular placement were run at 0.7 and 3.0v.All impedances were within normal limits.Follow up with the patient healthcare professional (hcp) indicated that the cause of the event was not determined and it was unknown if it was device related.X-rays and reprogramming was done.No intervention or outcome was provided regarding the event.Further follow-up is being conducted to obtain this information.
 
Manufacturer Narrative
Concomitant products: product id neu_ins_stimulator, serial# unknown, product type implantable neurostimulator; product id neu_ins_stimulator, serial# unknown, product type implantable neurostimulator; product id 3387s-40, lot# v679192, implanted: (b)(6) 2011, product type lead; product id 7482a51, serial# (b)(4), implanted: (b)(6) 2011, product type extension; product id 3387s-40, lot# v679192, implanted: (b)(6) 2011, product type lead; product id 7482a95, serial# (b)(4), implanted: (b)(6) 2011, product type extension; product id 7438, serial# (b)(4), product type programmer, patient; product id neu_ins_stimulator, serial# unknown, product type implantable neurostimulator.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
IMPLANTED SUBCORTICAL ELECTRICAL STIMULATOR (MOTOR DISORDERS)
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central avenue ne rcw215
minneapolis MN 55432
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key4218234
MDR Text Key5075371
Report Number3004209178-2014-20781
Device Sequence Number1
Product Code MRU
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H020007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/23/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/14/2014
Device Model Number37602
Device Catalogue Number37602
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/23/2014
Initial Date FDA Received10/31/2014
Date Device Manufactured03/21/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-