MAUDE Adverse Event Report: PHYSIO-CONTROL, INC LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR; DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR; DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES) Back to Search Results
Model Number 20
Device Problem Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 10/06/2014
Event Type  malfunction  
Manufacturer Narrative
Physio-control examined the removed system controller pcb assembly and determined that the cause of the reported issue was an integrated circuit (ic) chip, designator u61.The removed user interface pcb assembly was an ancillary part and there was no failure of this assembly.
 
Event Description
The customer contacted physio-control to report that their display would intermittently turn white and become non-responsive.A white display is indicative of a device lockup condition which could prevent defibrillation therapy from being delivered, if necessary.There was no patient use associated with the reported event.
 
Manufacturer Narrative
Physio-control examined the customer's device and verified the reported issue.Physio then replaced both the system controller (sc) and user interface (ui) pcb assemblies and after observing proper device operation through functional and performance testing the unit was returned to the customer for use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer Contact
jason march
4258674000
MDR Report Key4218456
MDR Text Key12620601
Report Number3015876-2014-01294
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063119
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 10/06/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number3202487
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/09/2014
Is the Reporter a Health Professional? Yes
Device Age8 YR
Event Location Hospital
Initial Date Manufacturer Received 12/15/2014
Initial Date FDA Received10/31/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/19/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/27/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse