FISHER & PAYKEL HEALTHCARE LTD BI-LEVEL/CPAP INSPIRATORY HEATED CIRCUIT; BTT Back to Search Results
Model Number RT319
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 10/08/2014
Event Type  malfunction  
Event Description
A hospital in (b)(6) reported to a fisher & paykel healthcare representative that the heater wire disconnect alarm sounded on an mr850 humidifier when an rt319 adult breathing circuit was connected.This was found before patient use.
 
Manufacturer Narrative
(b)(4).The product referred to in the event description is not sold in the usa but it is similar to a product which is sold in the usa.The 510(k) for that product is k983112.Method: the complaint rt319 adult breathing circuit was returned to fisher & paykel healthcare (b)(4) for evaluation.The heater wire was resistance tested.Results: the resistance test revealed that the heater wire was out of specification.Continuity testing and visual inspection revealed that the open circuit in the inspiratory heater wire was located between the heater wire and the right heater wire pin inside the overmoulded plug.A lot check revealed no other complaints of this type for the lot number provided.Conclusion: we are unable to determine what may have caused the observed damage on the heater wire.Resistance tests and visual inspections are performed on all breathing circuits during production and circuits that fail are rejected.This suggests that the heater wire became open circuit after the product was released for distribution.The user instructions that accompany the breathing circuit state the following: perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.Set appropriate ventilator alarms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BI-LEVEL/CPAP INSPIRATORY HEATED CIRCUIT
Type of Device
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
raymond yan
15365 barranca parkway
irvine, CA 92618-2216
9194534000
MDR Report Key4221430
MDR Text Key20655540
Report Number9611451-2014-00842
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/09/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRT319
Device Catalogue NumberRT319
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2014
Date Manufacturer Received10/09/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/11/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MR850 HUMIDIFIER
-
-