• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR; DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR; DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES) Back to Search Results
Model Number 20
Device Problem Failure To Run On AC/DC (1001)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/27/2014
Event Type  malfunction  
Event Description
The customer contacted physio-control to report that their device powered off twice when they tried to treat a patient.The device had been connected to a patient in cardiac arrest and it powered itself off.The device was restarted but powered off again.A backup device was then successfully used to provide defibrillation therapy to the patient.There was patient use associated with the reported event however, there was no adverse outcome for the patient reported.No patient details were provided.
 
Manufacturer Narrative
(b)(4): physio-control evaluated the device but could not verify the reported failure.Physio-control continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Manufacturer Narrative
Physio-control further evaluated the device but could not verify the reported failure.It was observed that an unrelated event code had been logged into the device's memory and that the main keypad was loose.The customer declined repair of the device and requested the device to be returned to them without being repaired.A cause of the reported failure could therefore not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer Contact
jason march
4258674000
MDR Report Key4225906
MDR Text Key4970521
Report Number3015876-2014-01309
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K063119
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 10/08/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number3202487
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2014
Is the Reporter a Health Professional? Yes
Device Age9 YR
Event Location Hospital
Initial Date Manufacturer Received 11/10/2014
Initial Date FDA Received11/05/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/26/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/07/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-