• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LIGASURE ADVANCE PISTOL GRIP; LIGASURE VESSEL SEALING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LIGASURE ADVANCE PISTOL GRIP; LIGASURE VESSEL SEALING SYSTEM Back to Search Results
Model Number LF5544
Device Problems Failure to Cut (2587); Material Protrusion/Extrusion (2979)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/15/2014
Event Type  malfunction  
Event Description
The customer reported that the device was unable to cut tissue.There was no injury to the patient.The device was returned to covidien and visual inspection discovered that the clear insulation was damaged and the webbing was protruding from the hinge.
 
Manufacturer Narrative
(b)(4).The clear insulation was damaged.The device failed hipot testing.The ifu states to prevent damage to the flexible insulation proximal to the jaws, confirm the handle is fully closed prior to insertion into and extraction from the cannula.Use the appropriately sized cannula to allow for easy insertion and extraction of the instrument.Failure to do so may impact the integrity of the flexible insulation.Cannulas with hard, non-beveled openings may cause the flexible insulation to retract, which may compromise the insulation.If retraction occurs, the instrument must be discarded.Do not attempt to clean the flexible insulation.Cleaning may damage insulation.The knife did not advance when the trigger was activated and it was found that the knife was hitting the back of the blue jaw seal plate.This can happen when the user places excessive tension on the jaws, forcing them out of alignment.The ifu cautions the user to not turn the rotation wheel when the handle is fully closed and latched.Product damage may occur.Do not apply force to the shaft of the instrument causing tension or bowing as this could make the knife difficult to deploy and the trigger may not return to its normal position.When the knife contacted the back of the seal plate, the trigger was forced and this caused the webbing to protrude.The webbing was sharp.The ifu states to gently pull the cutting trigger to engage the cutting mechanism.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE ADVANCE PISTOL GRIP
Type of Device
LIGASURE VESSEL SEALING SYSTEM
Manufacturer (Section D)
COVIDIEN LP
5920 longbow drive
boulder CO 80301
Manufacturer Contact
sharon murphy
5920 longbow drive
boulder, CO 80301
2034925267
MDR Report Key4226655
MDR Text Key16558710
Report Number1717344-2014-00969
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
Reporter Country CodeKS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 10/30/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2018
Device Model NumberLF5544
Device Catalogue NumberLF5544
Device Lot Number252805X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2014
Initial Date Manufacturer Received 10/30/2014
Initial Date FDA Received11/05/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/30/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-