Device Problem
Positioning Problem (3009)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/06/2014 |
Event Type
No Answer Provided
|
Event Description
|
The user facility reported their surgical table continued to tilt after the tilt button on the table was released.No injuries or procedural delays or cancellations were reported.
|
|
Manufacturer Narrative
|
The table laterally tilted left and continued after release of the button.The table is serviced by device technologies (b)(4) and was last serviced on (b)(4) 2014 and has been in use for over 10 years.The investigation of this event is currently in process.A follow-up report will be submitted when additional information becomes available.
|
|
Manufacturer Narrative
|
A service technician inspected the unit and identified that the table did not operate properly when running solely on battery power; no issues were noted when the table was plugged into a main power source.The technician replaced the table motor and control batteries.The table control board contains circuitry that will cause the table to shut down if the battery voltage is too low for correct circuit operation.Investigation by steris engineering determined the table's control board should be replaced as the control board may not have properly detected the battery voltage.The table control board was replaced and no additional issues have been reported.
|
|