HILL-ROM, INC. TOTALCARE BARIATRIC BED; A/C POWERED ADJUSTABLE HOSPITAL BED Back to Search Results
Model Number 1840
Device Problems Sticking (1597); Positioning Problem (3009)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/25/2014
Event Type  malfunction  
Event Description
Hill-rom received a report from a hill-rom technician stating the cpr would not flatten from the chair position.The bed was located in the hill-rom service center.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Manufacturer Narrative
The hill-rom technician found the sensors were out of calibration.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician calibrated the position sensors to resolve the issue.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOTALCARE BARIATRIC BED
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM, INC.
batesville IN
Manufacturer Contact
tony werner
1069 state route 46 east
batesville, IN 47006
8129312359
MDR Report Key4229916
MDR Text Key15817119
Report Number1824206-2014-02515
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122473
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 09/25/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/23/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number1840
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received09/25/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-