|
Device Problems
Device Operates Differently Than Expected (2913); Positioning Problem (3009)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/08/2014 |
Event Type
No Answer Provided
|
Event Description
|
On the (b)(6) because of intermittent diaphragm rotation inhibits, the service engineer changed and calibrated the coarse check and fine potentiometers.The calibration was checked and found to be ok.The calibration was also checked by the hospital physicist later on in the evening and found to be ok.No problems occurred until (b)(6) 2014 when the customer reported that the head angle was ok in the morning, but a shift of 7° occurred in the afternoon.Verification via remote phone call found that coarse and check potentiometers were still at 5.00 v at a head angle of 0°.The service engineer lead the customer through the calibration procedure, and after the calibration everything seemed to be normal again.
|
|
Manufacturer Narrative
|
The manufacturer's analysis has found that on (b)(6) (a.M.) the online matching with drrs (digitally reconstructed radiograph) showed the last patient before lunch was setup correctly with no positioning error.On (b)(6) (p.M.) the customer reported visual error with the collimator alignment.The projected field was not parallel to the imaging panel edges as it should have been.The manufacturer has not identified a root cause for this issue.Based on the information that has been provided the manufacturer concludes that this error was caused by a mechanical fixation not correctly carried out by the service engineer whilst completing maintenance upon the collimator assembly.The service engineer maintains that the assembly was fitted correctly but with the sequence of events: that the system was correct in the morning, then misaligned in the afternoon and subsequently resolved by carrying out a calibration.This series of events can be explained if the assembly was not correctly fitted i.E.The fine potentiometer assembly was not fully screwed together, and through slipping the assembly rotated causing the fault.A subsequent re-calibration of the collimator fine potentiometer after this scenario would cause the issue to be resolved.All fixings have now been inspected on site and the fine potentiometer replaced and re-calibrated.The fault has not reoccurred.The manufacturer will be carrying out a mandatory field safety corrective action for another issue relating to potentiometer assembly, this will involve a mandatory training course for all service engineers to reinforce the correct fixation method and the potential consequences of not correctly fitting these assemblies.The time schedule for this action is end of december 2014.
|
|
Search Alerts/Recalls
|
|
|