MAUDE Adverse Event Report: ETHICON INC. GYNECARE TVT ABREVO CONTINENCE SYSTEM; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. GYNECARE TVT ABREVO CONTINENCE SYSTEM; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC Back to Search Results
Catalog Number TVTOML
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Event Description
It was reported that a patient underwent a sling procedure in (b)(6) 2014.During the procedure, a portion of the introducer broke upon insertion of the mesh.The device was removed and another like device was used to complete the procedure with no adverse patient consequences.
 
Manufacturer Narrative
(b)(4).Conclusion: to date, the device has not been returned.If the product is returned for evaluation, any further info derived from the evaluation will be submitted in a supplemental 3500a form.In addition, a review of the batch manufacturing records was conducted and the batch met all finished goods release criteria.
 
Manufacturer Narrative
Conclusion: the actual device with 2 guides and the mesh protected by the plastic sheath attached to the two plastic needles was returned for evaluation.The device was visually evaluated.Upon evaluation, one needle was broken.The break was located at the level of the introduction hole.A portion of the needle was missing.The product was inspected and was conforming to specifications at release.Based on the evaluation, this is not a manufacturing issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GYNECARE TVT ABREVO CONTINENCE SYSTEM
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 015
Manufacturer (Section G)
ETHICON SARL-NEUCHATEL
puits-godet 20 2000 neuchatel
neuchatel
SZ  
Manufacturer Contact
guillermo villa
route 22 west po box 151
somerville, NJ 08876
9082180707
MDR Report Key4245856
MDR Text Key5016886
Report Number2210968-2014-15680
Device Sequence Number1
Product Code OTN
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K100936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/29/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2015
Device Catalogue NumberTVTOML
Device Lot Number3791970
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2014
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/16/2014
Initial Date FDA Received11/12/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/06/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/11/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-