• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER COULTER LH 750 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER COULTER LH 750 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL Back to Search Results
Catalog Number 6605632
Device Problems Thermal Decomposition of Device (1071); Smoking (1585)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/14/2014
Event Type  malfunction  
Event Description
The customer reported smoke and burning smell from the compressor of the coulter lh 750 hematology analyzer.There was no report of arcs, sparks, flames, or injuries in connection with this event.No patient results were impacted.The fire department was not called.A beckman coulter (bec) field service engineer (fse) was dispatched to evaluate the instrument.
 
Manufacturer Narrative
The fse confirmed the burning smell was caused by a worn plug connection at the rear of the pneumatic power supply module.The fse replaced the pneumatic power supply input one and the main power supply cable to resolve the issue.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER LH 750 HEMATOLOGY ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s kraemer blvd
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196
Manufacturer Contact
dung nguyen
250 s. kraemer blvd.
brea, CA 92821
7149614941
MDR Report Key4247155
MDR Text Key5015950
Report Number1061932-2014-02762
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K061574
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/14/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/13/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6605632
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/14/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-