PROMED INSTRUMENTS GMBH DORO (R) TEFLON SKULL CLAMP; HEADHOLDER Back to Search Results
Model Number 3003-009
Device Problem Difficult or Delayed Positioning (1157)
Patient Problem Laceration(s) (1946)
Event Date 12/12/2012
Event Type  malfunction  
Event Description
The customer reported on (b)(6) 2012 that there was a pt slippage on (b)(6) 2012 with laceration but no serious pt injury occurred.Surgery was completed.Pt positioning: prone."dr stated he checked for lock 3 times, when placed on pt it slipped and lacerated at left anterior scalp approx 5 cm laceration." the device was received at pmi for inspection on (b)(6) 2013.
 
Manufacturer Narrative
Root cause: device was not regularly inspected as part of the annual inspection program.Index knob evaluated and alleged slippage failure could not be duplicated.Index knob was stiff due to possible debris/residue.Pin pressure testing showed a deviation which was minimal.This did not contribute to the event.Suspected root cause: wrong pinning technique.Corrective action: instructed customer to properly position the pt's head.Ensure that the 2 skull pins in the rocker arm are equidistant from the centerline of the head and that all 3 pins are penetrating the skull equally.The angle of the pins should be as close to 90 degrees possible to the pt's skull.Test that the skull clamp ext.Arm is fully engaged in the clamp base.Lock the index knob so that the arrow and "lock" are aligned.See instruction manual, device was repaired and can be used as intended.Returned product and another copy of the instruction manual to customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORO (R) TEFLON SKULL CLAMP
Type of Device
HEADHOLDER
Manufacturer (Section D)
PROMED INSTRUMENTS GMBH
boetzinger strasse 38
freiburg 7911 1
GM  79111
Manufacturer Contact
lutz babilon
boetzinger strasse 38
freiburg 79111
GM   79111
6138422242
MDR Report Key4249930
MDR Text Key5013031
Report Number3003923584-2014-00008
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001808
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Nurse
Type of Report Initial
Report Date 01/18/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3003-009
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/02/2013
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/02/2013
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age58 YR
Patient Weight77
-
-