Model Number 1900
Device Problem Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/10/2014
Event Type  malfunction  
Event Description
Hill-rom received a report from a hill-rom technician stating the head section would not lower when the cpr lever was negated.The bed was located at the account in (b)(6).There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Manufacturer Narrative
The hill-rom technician found the cpr valve assembly stuck.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2007-2014.It is unknown if the facility performed any other preventative maintenance on this bed.The hill-rom technician replaced the cpr valve to resolve the issue.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOTALCARE BED
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM, INC.
batesville IN
Manufacturer Contact
john cummings
1069 state route 46 east
batesville, IN 47006
8129312869
MDR Report Key4250246
MDR Text Key18797213
Report Number1824206-2014-02591
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122473
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/10/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number1900
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received10/10/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-