HILL-ROM, INC. TOTALCARE BED; A/C POWERED ADJUSTABLE HOSP BED Back to Search Results
Model Number 1900
Device Problem Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/16/2014
Event Type  malfunction  
Event Description
Hill-rom received a report from the account stating the head section would not lower when cpr lever was used.The bed was located in engineering at the account.There was no pt/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Manufacturer Narrative
The hill-rom technician found the cpr valve was stuck in the manifold.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2013 and 2014.It is unk if the facility performed any other preventative maintenance on this bed.The tech replaced the cpr valve to resolve the issue.Based on this info, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOTALCARE BED
Type of Device
A/C POWERED ADJUSTABLE HOSP BED
Manufacturer (Section D)
HILL-ROM, INC.
batesville IN
Manufacturer Contact
tony werner
1069 state route 46 e
batesville, IN 47006
8129312359
MDR Report Key4259300
MDR Text Key5049875
Report Number1824206-2014-02614
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K962942
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Unknown
Type of Report Initial
Report Date 10/16/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number1900
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/16/2014
Initial Date FDA Received11/12/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-