The patient was undergoing a coil embolization procedure in an arteriovenous fistula using a penumbra coil 400.During the procedure, the physician attempted to reposition the penumbra coil 400; however, it unintentionally detached inside the px slim delivery microcatheter.The physician removed the penumbra coil 400 and the px slim delivery microcatheter as a system.There was no report of an adverse effect on the patient.
|
Result: the penumbra coil 400 was detached from the pusher assembly.The pet-lock was broken on the proximal end of the pusher assembly.These observations are consistent with a properly detached coil.The pull wire is no longer in the capture feature in the distal detachment tip (ddt).Conclusion: the complaint has been evaluated.The complaint indicates that the penumbra coil 400 unintentionally detached inside the pxslim while attempting to reposition the coil.Evaluation of the returned product confirmed that the penumbra coil 400 was detached from the pusher assembly however; the returned device does not show evidence of an unintentional detachment.The pet-lock was broken and retracted on the proximal end of the pusher assembly, which is consistent with a properly detached coil.The coil was intact and the pusher assembly was not damaged.Therefore, the root cause of the complaint could not be determined as the observations made during the device investigation are not consistent with the description of the event.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|