Upon investigation of the delta check failure for patient sample results generated with advia centaur cp br (ca 27.29 assay), the customer noted that there was excess calibrator left in the calibrator vials used to perform several calibrations of the assay.Customers are instructed to reconstitute calibrator g with 2 ml of di water.The customer feels that the calibrators were reconstituted with 5 mls of di water instead of 2 mls.Once the fresh vials of calibrators were reconstituted with the correct volume of di water, quality control results were within range.The system is performing within specification.No further investigation is required.The limitation section of the instructions for use states: "warning: do not use the advia centaur cp br assay as a screening test or for diagnosis.Normal levels of ca 27.29 do not always preclude the presence of disease.Note: do not interpret levels of ca 27.29 as absolute evidence of the presence or the absence of malignant disease.Before treatment, patients with confirmed breast carcinoma frequently have levels of ca 27.29 within the range observed in healthy individuals.Additionally, elevated levels of ca 27.29 can be observed in patients with nonmalignant diseases.Measurements of ca 27.29 should always be used in conjunction with other diagnostic procedures, including information the patient's clinical evaluation.".
|