The device history record (dhr) was reviewed and no deviations related to this failure mode were found.There are no non-conformances related to the reported issue.The product sample consisted of two adapters.One adapter corresponds to the arterial adult adapter and the other to the venous adapter.After a visual inspection, a crack on the thread region of the adapters was identified.As per the instructions for use, it is necessary to perform a visual inspection before using the device.Do not use the catheter if it appears damaged or defective.Overtightening catheter connections can crack the adapters.The device was more likely damaged during use.The most probable root cause can be due to over tightening of the adapter.A corrective action is not warranted at this time.It must be noted that in-process controls (such as personnel training, incoming quality acceptance testing for raw material, 100% in process visual inspection and visual acceptance sampling) are in place to prevent non-conforming product from leaving the manufacturing operations.As per procedure, manufacturing performs 100% visual inspection during production, which would identify cracked adapters in the catheter assembly.This complaint will be used for tracking and trending purposes.
|