• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COSTA RICA PALINDROME 19/36 KIT W/ SLOT; DIALYSIS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COSTA RICA PALINDROME 19/36 KIT W/ SLOT; DIALYSIS CATHETER Back to Search Results
Model Number 8888145014
Device Problem Crack (1135)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/05/2014
Event Type  malfunction  
Event Description
It was reported to covidien on (b)(6) 2014 that a customer had an issue with a dialysis catheter.The customer states that on (b)(6) 2014, the catheter body was found to be cracked.The catheter was pulled and replaced with a new one.
 
Manufacturer Narrative
Submit date: 11/25/2014.An investigation is currently underway.Upon completion, the results will be forwarded.
 
Manufacturer Narrative
The device history record was reviewed indicating that product was released accomplishing all quality standard requirements.There were no non-conforming issues reported during the production of this product for a similar condition as reported in this complaint.The complaint sample was not returned to the manufacturing site for review.Without the sample, it is not possible to determine a confirmed root cause of this issue.Should the sample be returned in the future, this complaint will be re-opened for further investigation.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.It must be noted that in-process controls (such as personnel training, incoming quality acceptance testing for raw material, 100% in process visual inspection and visual acceptance sampling) are in place to prevent nonconforming product from leaving the manufacturing operations.A corrective action is not warranted at this time.This information will be utilized for tracking and trending purposes to determine the need for corrective action.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PALINDROME 19/36 KIT W/ SLOT
Type of Device
DIALYSIS CATHETER
Manufacturer (Section D)
COSTA RICA
covidien manufacturing solulfons sa
edificio 820 calle #2 zona franca coyol
alajuela
CS 
Manufacturer (Section G)
COVIDIEN
covidien manufacturing solulfons sa
edificio 820 calle #2 zona franca coyol
alajuela
CS  
Manufacturer Contact
thom mcnamara
15 hampshire street
mansfield, MA 02048
5084524811
MDR Report Key4276746
MDR Text Key5296004
Report Number3009211636-2014-00180
Device Sequence Number1
Product Code MSD
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/03/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/25/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8888145014
Device Catalogue Number8888145014
Device Lot Number227828X
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received05/15/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/11/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-