AV-TEMECULA-CT MITRACLIP SYSTEM STEERABLE GUIDE CATHETER Back to Search Results
Catalog Number SGC01ST
Device Problem Torn Material (3024)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/04/2014
Event Type  malfunction  
Event Description
This report is being filed for the tear noted on the soft tip of the steerable guide catheter.A torn soft tip has the remote potential to cause serious injury if a piece of the soft tip is left behind in the anatomy.It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4.The steerable guiding catheter (sgc) and the clip delivery system (cds) were advanced into the anatomy; however, due to a low septal puncture, the decision was made to remove the mitraclip system to puncture the septum again in a higher position.An attempt was made to retract the cds into the sgc by pulling only the sleeve handle; however, resistance was noted.The clip was advanced and rotated with the dc handle many times under fluoroscopic guidance.After several attempts, in addition to using - deflection to close the clip more, it was possible to retract the cds into the sgc.The physician did not feel that the cds was stuck with the sgc tip, but only met resistance with it.No force was applied.After sgc and cds were removed, a tear was noted on the sgc soft tip and a bent gripper was observed on the cds.No portion of the sgc soft tip was missing.After the second puncture, a new cds and sgc were used.Two mitraclips were implanted successfully and the mr was reduced to 1.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
(b)(4).The incident information was reviewed; however, a failure analysis of the complaint device could not be performed because the product was not returned for analysis.Therefore, the analysis of this complaint will be an assessment of the manufacturing records/complaint history and information provided to abbott vascular.Potential causes for difficulty retracting the clip delivery system (cds) into the guide tip, resulting in soft tip tear/damage can be caused by, but are not limited to, manufacturing anomalies (inner diameter [id] of the tip not within specification), user technique or procedural conditions (curves on the guide during cds removal).With respect to the procedural conditions and/or user technique, clip getting caught on guide resulting in soft tip damage can be influenced by the clip not being fully closed upon removal, the orientation of the clip with respect to the guide tip or curves on the sgc applied by the user.Based on the information reviewed, the reported difficult to remove cds resulting in soft tip damage appears to be related to procedural conditions / user technique and not a product quality deficiency.A review of the device history record revealed no non-conformances for the reported lot.Additionally, a review of the complaint-handling database identified no other reported incidents for torn material from this lot.Based on the information reviewed, there is no indication of a product deficiency.The clip delivery system (cds) referenced is filed under a separate medwatch report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP SYSTEM STEERABLE GUIDE CATHETER
Type of Device
STEERABLE GUIDE CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer (Section G)
MENLO PARK, CA REG# 3005070406
abbott vascular
26531 ynez rd.
temecula CA 92591 462
Manufacturer Contact
connie speck
abbott vascular
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key4277288
MDR Text Key5010201
Report Number2024168-2014-07764
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K112239
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 11/04/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/25/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2015
Device Catalogue NumberSGC01ST
Device Lot Number10364185
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/04/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MITRACLIP SYSTEM, CLIP DELIVERY SYSTEM -(B)(4)
-
-