ST. JUDE MEDICAL - NEUROMODULATION GENESIS PULSE GENERATOR, 8-CHANNEL; SCS IPG Back to Search Results
Model Number 3608
Device Problem Premature End-of-Life Indicator (1480)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/04/2014
Event Type  Injury  
Event Description
It was reported the patient's (finland) ipg displayed a low battery warning due to possible premature battery depletion, however, the patient did not lose stimulation.Surgical intervention was undertaken and the ipg was explanted and replaced.The patient reported effective stimulation coverage postoperative.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Manufacturer Narrative
Evaluation codes results: a visual inspection of the ipg did not reveal any discrepancies.Communication with the ipg as received was normal.The device was subjected to a functional test on automated test equipment (ate).This tester verifies the electrical performance of the control/communication circuitry, and output signal integrity.All portions of ate testing passed.The pulse load testing of the ipg battery confirmed the ipg low battery indication noted in the field was due to ipg battery passivation.The ipg battery was more prone to passivation at the settings used during the device implant period.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GENESIS PULSE GENERATOR, 8-CHANNEL
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key4279823
MDR Text Key5030004
Report Number1627487-2014-15806
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeFI
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/10/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date04/28/2016
Device Model Number3608
Device Lot Number4548085
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/10/2015
Initial Date FDA Received11/26/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/06/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/03/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age52 YR
-
-