MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problems Air Leak (1008); Defibrillation/Stimulation Problem (1573)
Patient Problems Low Blood Pressure/ Hypotension (1914); ST Segment Elevation (2059); Ventricular Fibrillation (2130)
Event Date 11/06/2014
Event Type  Injury  
Event Description
During a cryoablation procedure, there was a drop in the patient's blood pressure.The physician noticed air bubbles in the line that connects the sheath to the heparinized saline bag; the nurse had just changed the flush bag to a new bag.St elevation occurred shortly after the air bubbles were noticed.Air embolism reported to have caused the st elevation which caused ventricular fibrillation (vf).Patient could not be shocked out of vf.Chest compressions were performed until patient stabilized (about 9 minutes).St elevation subsided and patient went into normal sinus rhythm.The cryoablation procedure was aborted.
 
Manufacturer Narrative
The device was not returned for investigation.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
16771 chemin ste-marie
kirkland,qc H9H 5 H3
CA  H9H 5H3
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
16771 chemin ste-marie
kirkland,qc H9H 5 H3
CA   H9H 5H3
Manufacturer Contact
voula radiotis
16771 chemin ste-marie
kirkland,qc H9H 5-H3
CA   H9H 5H3
5146941212
MDR Report Key4286599
MDR Text Key16181320
Report Number3002648230-2014-00207
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative,company representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/06/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4FC12
Device Catalogue Number4FC12
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/06/2014
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-