• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - MALTA; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MALTA; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number NGB8064M
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Event Description
It was reported that a reconstitution device leaked.This occurred during reconstitution.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
(b)(4).Evaluation summary: the actual sample was not available for evaluation; however, an unused companion sample was received.Visual inspection of the companion sample did not identify any abnormalities that could have contributed to the reported condition.Simulated use testing was performed by connecting the sample to a vial and a viaflo bag.Reconstitution was performed without any issues or leaks noted.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE - MALTA
a47 industrial estate
malta b
marsa
MT 
Manufacturer (Section G)
BAXTER HEALTHCARE - MALTA
a47 industrial estate
malta b
marsa
MT  
Manufacturer Contact
christina arnt
25212 w. illinois route 120
round lake, IL 60073
2242703198
MDR Report Key4288145
MDR Text Key5193222
Report Number1416980-2014-43325
Device Sequence Number1
Product Code LHI
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 11/04/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date07/31/2018
Device Catalogue NumberNGB8064M
Device Lot Number13H01V510
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/09/2014
Initial Date FDA Received12/01/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/06/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/02/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-