Investigation results for the returned platform as follows: visual inspection was performed and the customer's reported complaint of the lifeband falling off of the platform's channel die cast assembly was confirmed.During functional testing, the platform ran for 20 minutes using a test mannequin and an additional 10 minutes using a large resuscitation test fixture (lrtf) equivalent to a 250 pound patient with no problems encountered.A review of the platform's archive was performed and found that there were no user advisories observed on the reported event date of (b)(6) 2014.Based on the investigation, the part identified for replacement was the channel die cast assembly.In summary, the reported complaint was confirmed during visual inspection and attributed to the platform's channel die cast assembly being wider than intended.
|