• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problems Component Falling (1105); Fitting Problem (2183)
Patient Problem No Patient Involvement (2645)
Event Date 11/12/2014
Event Type  malfunction  
Event Description
It was reported that during a device check by the distributor, the autopulse lifeband fell out of the autopulse platform.The distributor reported that the distance between the black tabs on the platform that hold the lifeband in place is too wide.They tried different lifebands but that did not resolve the issue.No patient involvement was reported.No further information was provided.
 
Manufacturer Narrative
The autopulse platform in complaint was returned to zoll on 12/02/2014 for investigation.However, investigation is still in progress.A supplemental report will be filed when investigation has been completed.
 
Manufacturer Narrative
Investigation results for the returned platform as follows: visual inspection was performed and the customer's reported complaint of the lifeband falling off of the platform's channel die cast assembly was confirmed.The platform was functionally tested and passed all test requirements.The platform ran for 15 minutes using a large resuscitation test fixture (lrtf) with no problems encountered.A review of the platform's archive was performed and found that there were no user advisories noted on the reported event date of (b)(6) 2014.Based on the investigation, the part identified for replacement was the channel die cast assembly.In summary, the reported complaint was confirmed during visual inspection and attributed to the platform's channel die cast assembly being wider than intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
joy patel
2000 ringwood ave,
san jose, CA 95131
4084192957
MDR Report Key4295789
MDR Text Key5195499
Report Number3010617000-2014-00644
Device Sequence Number1
Product Code DRM
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/12/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/03/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberMODEL 100
Device Catalogue Number8700-0740-11
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received12/04/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/27/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-