• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION OCTRODE LEAD KIT, 60CM LENGTH; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION OCTRODE LEAD KIT, 60CM LENGTH; SCS LEAD Back to Search Results
Model Number 3186
Device Problems Device Operates Differently Than Expected (2913); Unintended Movement (3026)
Patient Problems Fall (1848); Inadequate Pain Relief (2388)
Event Date 11/03/2014
Event Type  Injury  
Event Description
It was reported the patient's (united kingdom) scs system is not functioning properly since the patient fell.X-ray images indicated the lead has moved and there is a kink and coiling of the lead at the distal end of the anchor.Surgical intervention is planned for a later date to address the issue.
 
Event Description
Additional information received identified the patient's scs lead was explanted and replaced with a new one.Reportedly, effective stimulation therapy was restored.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE LEAD KIT, 60CM LENGTH
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key4301251
MDR Text Key5270281
Report Number1627487-2014-02832
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/14/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Expiration Date03/28/2012
Device Model Number3186
Device Lot Number3088700
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/14/2015
Initial Date FDA Received12/05/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/11/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/23/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 3716, SCS IPG; MODEL 1192, SCS ANCHOR
Patient Outcome(s) Other;
-
-