• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PKR INSERT X3 #2 RM/LL -8MM; IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PKR INSERT X3 #2 RM/LL -8MM; IMPLANT Back to Search Results
Catalog Number 5630-G-228
Device Problems Difficult to Insert (1316); Material Distortion (2977); Material Integrity Problem (2978)
Patient Problems Injury (2348); No Known Impact Or Consequence To Patient (2692)
Event Date 11/10/2014
Event Type  malfunction  
Event Description
It was reported that 8mm#2 insert could not be inserted again and again.Then, it was broken during medical procedure.A 9mm #2 insert could not be inserted, so #2 tibial component was removed.And, tibia was recut and #1 tibial component was implanted instead of #2 tibial component.The surgeon controlled the gap and cutting area of bone, but it was difficult to inserted and finally inserted.
 
Manufacturer Narrative
When completed, the investigation results will be submitted in a supplemental report.
 
Manufacturer Narrative
An event regarding difficulty seating a triathlon pkr insert into a baseplate was reported.The event was confirmed.Method adn results: device evaluation and results: a visual inspection showed the anterior portion of the locking feature was damaged, which would prevent it from fully seating.Medical records received and evaluation: not performed as patient factors did not contribute to the event.Device history review: all devices accepted into final stock conformed to specification.Complaint history review: there have been no similar reported events for this lot id.Conclusions: the investigation concluded that the observed damage on the anterior portion of the locking feature was preventing the device from being fully seated.This damage likely occurred as a result of attempting to implant the insert while it was not fully aligned with the baseplate.
 
Event Description
It was reported that 8mm#2 insert could not be inserted again and again.Then, it was broken during medical procedure.9mm #2 insert could not be inserted, so #2 tibial component was removed.And, tibia was recut and #1 tibial component was implanted instead of #2 tibial component.The surgeon controlled the gap and cutting area of bone, but it was difficult to inserted and finally inserted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PKR INSERT X3 #2 RM/LL -8MM
Type of Device
IMPLANT
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
beverly lima
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key4303377
MDR Text Key5193566
Report Number0002249697-2014-04523
Device Sequence Number1
Product Code HRY
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K082567
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,health professional,oth
Reporter Occupation Other
Type of Report Initial
Report Date 11/10/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2019
Device Catalogue Number5630-G-228
Device Lot NumberMNDV2P
Other Device ID NumberSTERILE LOT# MSGND11H7
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/24/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/10/2014
Initial Date FDA Received12/08/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/12/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Weight63
-
-