• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM INC. VERSACARE BED; A/C POWERED ADJUSTABLE HOSPITAL BED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM INC. VERSACARE BED; A/C POWERED ADJUSTABLE HOSPITAL BED Back to Search Results
Model Number 3200
Device Problem Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/02/2014
Event Type  malfunction  
Event Description
Hill-rom received a report from the account stating the brakes were not holding.The bed was located at the account.There was no patient/ user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Manufacturer Narrative
The hill-rom technician found the head brake casters inoperable.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2009-2014.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the casters to resolve the issue.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACARE BED
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM INC.
batesville IN
Manufacturer Contact
john cummings
1069 state route 46 east
batesville, IN 47006
8129312869
MDR Report Key4304515
MDR Text Key15317539
Report Number1824206-2014-02716
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Unknown
Type of Report Initial
Report Date 11/02/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/25/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number3200
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received11/02/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-